Quantikine Ivd Stfr Enzyme Linked Immunosorbent Assay (Elisa)
Immunochemical, Transferrin. 510(k) K970718, from R&D Systems, Inc. (Minneapolis, MN). Product code JNM, class 2.
As of Nov 17, 2026, K970718 is an FDA 510(k) held by R&D Systems, Inc., substantially equivalent on May 27, 1997 after 90 days, with 3 recalls tied to it in FDA's recall database.
- Substantially equivalent: May 27, 1997
- Review time: 90 days
- Recalls: 3
- Cited as a predicate by: 4 devices
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Access Ferritin AssayK926221 · Bio-Rad Laboratories, Inc. · Jan 13, 1993
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- Nichols Advantage Chemiluminescence Soluble Transferrin Receptor Immunoassay, Model…K992188 · Nichols Institute Diagnostics · Dec 6, 1999
- N Latex Stfr ReagentK991157 · Dade Behring, Inc. · Aug 2, 1999
- TfrK981208 · Ramco Laboratories, Inc. · Oct 21, 1998
- Access Stfr, Access Stfr Calibrators, And Access Stfr Qc With Models, A32493, A32494, And…K080634 · Beckman Coulter, Inc. · Aug 11, 2008
Recalls
Recalls FDA ties to K970718, newest first; each opens FDA's recall record| Jan 20, 2021 | Z-1255-2021R&D Systems, Inc. received two complaints that the standard signal was low and controls were not within the specification range provided with the kit. Investigation confirmed low standard signal and controls running out… | Terminated |
| Nov 1, 2018 | Z-0691-2019The Human sTfR Quantikine IVD ELISA Kit, catalog # DTFR1, lot P180151 was packed with the incorrect control data card number 750228.4. | Terminated |
| Aug 24, 2016 | Z-2887-2016R&D Systems, Inc. received two customer complaints that results of the Controls provided with the Quantikine IVD Human sTfR Immunoassay kit were out of the range high. Internal testing within R&D Systems confirmed a… | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Feb 26, 1997
- Review panel
- Immunology
- Regulation
- 21 CFR 866.5880
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files