Disposable Transrectal/Transvaginal/Endocavity Ultrasound Needle Instrument Guide

Transducer, Ultrasonic, Diagnostic. 510(k) K970514, from CIVCO Medical Instruments Co., Inc. (Kalona, IA). Product code ITX, class 2.

As of Nov 19, 2026, K970514 is an FDA 510(k) held by CIVCO Medical Instruments Co., Inc., substantially equivalent on Jun 20, 1997 after 129 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 129 days
  • Recalls: 3
  • Cited as a predicate by: 6 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K970514, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 30, 2018Z-2947-2018, Z-2946-2018The assembly of the needle guides were found to be either missing adhesive or did not have adequate adhesive to hold the assembled cannula in place on the needle guide.Terminated
Jan 16, 2008Z-1054-2008Incorrect Needle Guide- The individual pouches contained V53W Endocavity Needle guide(p/n/034-054) instead of the Shimadzu T/R Needle guide (p/n134-054).Terminated

Other ITX clearances

The newest 510(k)s for the same product code; their median review was 178 days; see every ITX device

Record

Type
Traditional
Received
Feb 11, 1997
Review panel
Radiology
Regulation
21 CFR 892.1570

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files