Arca
Skip to content
Tools
Tools

System 96 Intra-Aortic Balloon Pump (Iabp)

System, Balloon, Intra-Aortic And Control. 510(k) K965236, from Datascope Corp. (Paramus, NJ). Product code DSP, class 2.

As of Nov 26, 2026, K965236 is an FDA 510(k) held by Datascope Corp., substantially equivalent on Jul 1, 1997 after 182 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 182 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The following are the indications for use for the System 96 Intra Aortic Balloon Pump: o Refractory Ischemia o Myocardial Infarction o Preoperative Hemodynamic Stability o Weaning from Cardio Pulmonary Bypass o Refractory Ventricular Failure o Cardiogenic Shock o Unstable Refractory Angina o Septic Shock o Angioplasty Support o Intraoperative Pulsatile Flow Generation Any other uses are not…

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K965236, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 21, 2014Z-1523-2014Potential mechanical failure of the fan assembly associated with the power supply. A fan assembly failure could result in the power supply overheating and cause the IABP to shut down without warning.Terminated
Mar 16, 2011Z-2414-2011Several models of the Intra-Aortic Balloon Pumps are being recalled due to a potential problem with the fan's assembly which may cause the power supply to detect an overheating event and shut down.Terminated

Other DSP clearances

The newest 510(k)s for the same product code; see every DSP device

Record

Type
Traditional
Received
Dec 31, 1996
Review panel
Cardiovascular
Regulation
21 CFR 870.3535

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files