Arca
Skip to content
Tools
Tools

Edge/Vista System Enhancement Package

System, Nuclear Magnetic Resonance Imaging. 510(k) K964626, from Philips Medical Systems (Cleveland), Inc. (Highland Heights, OH). Product code LNH, class 2.

As of Nov 18, 2026, K964626 is an FDA 510(k) held by Philips Medical Systems (Cleveland), Inc., substantially equivalent on Feb 10, 1997 after 83 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 83 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K964626, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 7, 2013Z-0033-2014There is a potential for water to collect in the vent pipe elbow in the magnet venting system. Water may freeze, blocking the venting system.Terminated
Sep 19, 2008Z-2431-2010, Z-2429-2010, Z-2430-2010Water may collect and freeze in the vent pipe. This can rupture the magnet vessel during a cryogen transfer, and release helium into the MRI suite.Terminated

Other LNH clearances

The newest 510(k)s for the same product code; their median review was 143 days; see every LNH device

Record

Type
Traditional
Received
Nov 19, 1996
Review panel
Radiology
Regulation
21 CFR 892.1000

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.