Edge/Vista System Enhancement Package
System, Nuclear Magnetic Resonance Imaging. 510(k) K964626, from Philips Medical Systems (Cleveland), Inc. (Highland Heights, OH). Product code LNH, class 2.
As of Nov 18, 2026, K964626 is an FDA 510(k) held by Philips Medical Systems (Cleveland), Inc., substantially equivalent on Feb 10, 1997 after 83 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 83 days
- Recalls: 4
- Cited as a predicate by: 3 devices
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- 1.5T Edge and 1.0T Vista SystemsK931544 · Philips Medical Systems (Cleveland), Inc. · Nov 22, 1993
- Epi-Ii System Option W/High Performance GradientsK954646 · Philips Medical Systems (Cleveland), Inc. · Jan 18, 1996
Cited as a predicate
3 later 510(k) summaries name this device as a predicate- Ipass Bolus TrackingK992412 · Philips Medical Systems (Cleveland), Inc. · Aug 17, 1999
- Outlook System Enhancement PackageK974844 · Philips Medical Systems (Cleveland), Inc. · Mar 10, 1998
- Asset ApolloK971884 · Philips Medical Systems (Cleveland), Inc. · Aug 20, 1997
Recalls
Recalls FDA ties to K964626, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Aug 7, 2013 | Z-0033-2014There is a potential for water to collect in the vent pipe elbow in the magnet venting system. Water may freeze, blocking the venting system. | Terminated |
| Sep 19, 2008 | Z-2431-2010, Z-2429-2010, Z-2430-2010Water may collect and freeze in the vent pipe. This can rupture the magnet vessel during a cryogen transfer, and release helium into the MRI suite. | Terminated |
Other LNH clearances
The newest 510(k)s for the same product code; their median review was 143 days; see every LNH device- MAGNETOM Free.XLK262011 · Siemens Healthineers AG · Nov 5, 2026
- ECHELON SynergyK262902 · Fujifilm Corporation · Oct 9, 2026
- Vista AI ScanK260092 · Vista AI, Inc. · Oct 7, 2026
- SwiftSight-BodyCompositionK261612 · Airs Medical, Inc. · Oct 1, 2026
- APERTO Lucent MRI SystemK261283 · Fujifilm Corporation · Sep 16, 2026
- Embrace Vision Neonatal MRI SystemK261758 · Aspect Imaging, Ltd. · Sep 16, 2026
- MAGNETOM Free.Max; MAGNETOM Free.StarK260758 · Siemens Shenzhen Magnetic Resonance , Ltd. · Sep 1, 2026
- Swoop® Portable MR Imaging® SystemK262622 · Hyperfine, Inc. · Aug 19, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 19, 1996
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1000
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files