Fast-Cath Transseptal Catheter Introducer

Introducer, Catheter. 510(k) K964518, from Daig Corp. (Minnetonka, MN). Product code DYB, class 2.

As of Nov 21, 2026, K964518 is an FDA 510(k) held by Daig Corp., substantially equivalent on May 14, 1997 after 201 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 201 days
  • Recalls: 4
  • Cited as a predicate by: 8 devices

Cited as a predicate

8 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K964518, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 28, 2012Z-0617-2013St. Jude Medical is recalling a single batch of Fast Cath Transseptal Guiding Introducers which was packaged as a 180 degree curve style introducer instead of a 90 degree curve style introducer as indicated on the…Terminated
Oct 5, 2006Z-0535-2007Limited number of FastCath Transseptal Catheter Introducers (REF #406850) are susceptible to the irrigation extension tube on the side port becoming disconnected when a pull force is applied.Terminated
Aug 4, 2004Z-1425-04, Z-1426-04St. Jude Medical received two complaints from customers involving the Telesheath Left Atrial Introducer System side arm detaching from the hub of the inner sheath. Subsequent testing of Telesheath inventory at St. Jude…Terminated

Other DYB clearances

The newest 510(k)s for the same product code; their median review was 71 days; see every DYB device

Record

Type
Traditional
Received
Oct 25, 1996
Review panel
Cardiovascular
Regulation
21 CFR 870.1340

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files