Bf 240/P240/IT240 Bronchovideoscope & Accessories
Bronchoscope (Flexible Or Rigid). 510(k) K963033, from Olympus America, Inc. (Melville, NY). Product code EOQ, class 2.
As of Nov 17, 2026, K963033 is an FDA 510(k) held by Olympus America, Inc., substantially equivalent on Sep 9, 1996 after 35 days, with 6 recalls tied to it in FDA's recall database.
- Substantially equivalent: Sep 9, 1996
- Review time: 35 days
- Recalls: 6
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe BF Type 240 Bronchovideoscopes have been specifically designed to be used with the recommended electrocautery devices for endoscopic diagnosis and treatment within the airways and tracheobronchial tree. *66* *K963033*
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Evis 200 SystemK931154 · Olympus Corp. · Oct 7, 1993
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- Olympus Lf-Tp And Lf-Dp Tracheal Intubation Fiberscopes, Accessories And Ancillary…K981543 · The Olympus Optical Co. · Jun 8, 1998
- Modification To Olympus Bronchoscopes BF-40 Series, BF-240 Series, And BF-160 SeriesK023984 · The Olympus Optical Co. · Jan 31, 2003
- Visera Hysterovideoscope Olympus Hyf Type VK022445 · Olympus Optical Co., Ltd. · Jan 24, 2003
Recalls
Recalls FDA ties to K963033, newest first; each opens FDA's recall record| Sep 25, 2023 | Z-0327-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for… | Open, Classified |
| Sep 25, 2023 | Z-0337-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for… | Open, Classified |
| Jul 27, 2023 | Z-2503-2023Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large. | Open, Classified |
| Aug 8, 2018 | Z-3286-2018The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port. | Terminated |
| Aug 8, 2018 | Z-3289-2018The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port. | Terminated |
| Aug 8, 2018 | Z-3298-2018The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port. | Terminated |
Other EOQ Clearances
The newest 510(k)s for the same product code; their median review was 229 days; see every EOQ device- Single-use Flexible BronchoscopeK261637 · Hunan Vathin Medical Instrument Co., Ltd. · Sep 23, 2026
- Ambu® aScope 5 Broncho 2.7/1.2; Ambu® aScope 5 Broncho 4.2/2.2; Ambu® aScope 5 Broncho…K260309 · Ambu A/S · Sep 16, 2026
- 3.8mm Slim ProVu Video Bronchoscope (090-02-0038U); 5.0mm Regular ProVu Video…K254026 · Flexicare Medical Limited. · Sep 2, 2026
- MONARCH Platform (MON-000008)K260382 · Auris Health, Inc. · Jul 25, 2026
- LungFlow Airway CatheterK254037 · Free Flow Medical, Inc. · Jul 10, 2026
- FUJIFILM Endoscope Model EB-840S; FUJIFILM Endoscope Model EB-840TK253137 · Fujifilm Corporation · Jun 16, 2026
- SmartClip Delivery Catheter (ADR-1715)K260565 · Elucent Medical · May 14, 2026
- Evis Exera Iii Bronchovideoscope (Olympus BF-Q190); Evis Exera Iii Bronchovideoscope…K252341 · Olympus Medical Systems Corp. · Apr 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 5, 1996
- Review panel
- Ear, Nose, Throat
- Regulation
- 21 CFR 874.4680
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files