Arca
Skip to content
Tools
Tools

Ecat Exact & Ecat Exact Hr+

System, Tomography, Computed, Emission. 510(k) K962797, from Cti Pet Systems, Inc. (Knoxville, TN). Product code KPS, class 2.

As of Nov 18, 2026, K962797 is an FDA 510(k) held by Cti Pet Systems, Inc., substantially equivalent on Oct 15, 1996 after 89 days, with 14 recalls tied to it in FDA's recall database.

  • Review time: 89 days
  • Recalls: 14
  • Cited as a predicate by: 3 devices

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K962797, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 23, 2007Z-0273-2008Incorrect Data Results - ECAT System's software versions 7.2.2 and 7.4 may calculate the decay correction for emission data incorrectly while performing whole-body studies with all transmissions followed by all…Terminated
Jul 23, 2007Z-0248-2008, Z-0250-2008, Z-0249-2008, Z-0252-2008, Z-0247-2008, Z-0246-2008, Z-0251-2008Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix…Terminated
Jul 23, 2007Z-0272-2008, Z-0270-2008, Z-0269-2008, Z-0271-2008Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all…Terminated
Apr 5, 2005Z-0819-05Failure to properly secure the Mobile Bed Supports during transit could cause the support screws on the Patient Handling System to become loose and possibly break due to increased vibration during transit.Terminated
Mar 4, 2003Z-0674-03Incorrect patient information is displayed in the reconstructed images of a whole body patient scan.Terminated

Other KPS clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every KPS device

Record

Type
Traditional
Received
Jul 18, 1996
Review panel
Radiology
Regulation
21 CFR 892.1200

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.