Arca
Skip to content
Tools
Tools

Coaguchek Pst System

Test, Time, Prothrombin. 510(k) K962571, from Boehringer Mannheim Corp. (Indianapolis, IN). Product code GJS, class 2.

As of Nov 18, 2026, K962571 is an FDA 510(k) held by Boehringer Mannheim Corp., substantially equivalent on Apr 22, 1997 after 295 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 295 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K962571, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 28, 2007Z-0737-2007The product was erroneously distributed to home users in contrast to its use for healthcare professionals only.Terminated
Oct 18, 2006Z-0227-2007Erroneous Test Results: Monitor may display 'error' message or report falsely elevated patient results caused by insufficient amounts of active ingredient (thromboplastin) in the test strips.Terminated
Mar 28, 2005Z-0098-06Breaking the glass ampoule without first taking precautions may result in cuts/user injury from the glass.Terminated
Sep 12, 2003Z-1293-03Erroneous patient results: may be falsely elevated or lowered, due to a packaging defect that allows air and moisture to enter the sealed pouch.Terminated

Other GJS clearances

The newest 510(k)s for the same product code; their median review was 299 days; see every GJS device

Record

Type
Traditional
Received
Jul 1, 1996
Review panel
Hematology
Regulation
21 CFR 864.7750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.