Continuflo Solution Set/Secondary Medication Set
Set, Administration, Intravascular. 510(k) K961225, from Baxter Healthcare Corp (Round Lake, IL). Product code FPA, class 2.
As of Nov 20, 2026, K961225 is an FDA 510(k) held by Baxter Healthcare Corp, substantially equivalent on Jun 21, 1996 after 84 days, with 7 recalls tied to it in FDA's recall database.
- Review time: 84 days
- Recalls: 7
- Cited as a predicate by: 3 devices
- Clearlink System Solution SetK180739 · Baxter Healthcare Corporation · May 28, 2019
- Solution Set/Extension SetK142011 · Baxter Healthcare Corporation · Aug 18, 2014
- Large Bore Stopcock With Rotating Male Luer Lock/Large Bore Stopcock With Extension Set, Gang Large Bore Stopcock ManifoK130245 · Baxter Healthcare Corp · Mar 1, 2013
| Dec 12, 2019 | Z-0794-2020, Z-0793-2020Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the… | Open |
| Dec 6, 2011 | Z-0475-2012During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a… | Terminated |
| Jul 21, 2011 | Z-2920-2011The Solution Sets are being recalled because the position of the slide clamp and roller clamp may have been reversed during assembly. The reversed orientation of the slide clamp and roller clamp will cause the set to be… | Terminated |
| Dec 15, 2005 | Z-0653-06The extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication. | Terminated |
| Feb 28, 2005 | Z-0845-05Possible occlusion of the Clearlink injection site valve due to inadequate silicone lubrication of the slit in the upper seal of the valve. | Terminated |
| Jul 16, 2003 | Z-1152-03There are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process. | Terminated |
- Gravity Blood Sets with 200µm FilterK260261 · B.Braun Medical, Inc. · Oct 22, 2026
- Ecomed Disposable Administration SetsK254123 · Ecomed Solutions, LLC · Sep 11, 2026
- iiSure Infusion SetK261530 · Deka Research and Development · Jun 3, 2026
- AMSafe Administration Set (BA-70075, BA-70071-D, BA-70070, BA-70071-P, BA-70072, BA-70072-F, BA-70071-DF,…K252543 · Amsino International, Inc. · May 4, 2026
- PuraCath Firefly Needleless Connector IT (9005)K251375 · Puracath Medical, Inc. · Feb 12, 2026
- nSet+ Stabilization Set [6426733]; nSet+ Stabilization Set with nSyte Needle Free Connector (NFC) [6426731]; nSyte…K251257 · Np Medical · Dec 4, 2025
- BD Alaris Pump Infusion SetK250325 · Carefusion (Bd) · Oct 30, 2025
- EZ™ IV Administration SetK251814 · Epic Medical Pte. , Ltd. · Aug 29, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Mar 29, 1996
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5440
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files