Arca
Skip to content
Tools
Tools

Nexgen Compl. Knee Sol. Legacy Posterior Stabilized/Constrained Condylar Knee Femoral Components/Articular Surfaces

Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer. 510(k) K960279, from Zimmer, Inc. (Warsaw, IN). Product code JWH, class 2.

As of Nov 18, 2026, K960279 is an FDA 510(k) held by Zimmer, Inc., substantially equivalent on Apr 26, 1996 after 98 days, with 21 recalls tied to it in FDA's recall database.

  • Review time: 98 days
  • Recalls: 21
  • Cited as a predicate by: 9 devices

Cited as a predicate

9 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K960279, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 6, 2025Z-1038-2025, Z-1039-2025It was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope.Open
Nov 29, 2017Z-1040-2018, Z-1034-2018, Z-1037-2018, Z-1035-2018, Z-1036-2018The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.Terminated
Jan 11, 2016Z-0834-2016LDPE bag containing the implant adheres to the highly polished implant surface.Terminated
Jan 11, 2016Z-1146-2016As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.Terminated
Apr 2, 2015Z-1575-2015The affected lot is being recalled because it was packaged without the locking screw. The locking screw packaged with the augment is used to fix the augment to the femoral component. This has the potential to delay…Terminated
Oct 11, 2013Z-1023-2014Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations…Terminated
Sep 15, 2010Z-0535-2011, Z-0532-2011, Z-0530-2011, Z-0536-2011, Z-0528-2011, Z-0531-2011, Z-0533-2011, Z-0529-2011, Z-0534-2011These lots of Zimmer NexGen Complete Knee Solution LPS Femoral Component and Zimmer NexGen Complete Knee Solution LPS Flex Gender Femoral Component exhibit a nonconforming internal CAM radius. The condition is…Terminated
Apr 16, 2010Z-0609-2011The impactor may disassemble or detach during use. The detached piece may not be observed intraoperatively and may be left in the surgical site. Complications may include pain, infection, and prolongation in surgery.…Terminated

Other JWH clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every JWH device

Record

Type
Traditional
Received
Jan 19, 1996
Review panel
Orthopedic
Regulation
21 CFR 888.3560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.