Arca
Skip to content
Tools
Tools

Trimport Implantable Access System (Modification)

Port & Catheter, Implanted, Subcutaneous, Intravascular. 510(k) K954258, from Gerard Medical Enterprises, Inc. (Charlton, MA). Product code LJT, class 2.

As of Nov 22, 2026, K954258 is an FDA 510(k) held by Gerard Medical Enterprises, Inc., substantially equivalent on May 24, 1996 after 262 days, with 12 recalls tied to it in FDA's recall database.

  • Review time: 262 days
  • Recalls: 12

Recalls

Recalls FDA ties to K954258, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 13, 2010Z-1883-2010, Z-1881-2010, Z-1886-2010, Z-1882-2010, Z-1887-2010, Z-1884-2010, Z-1885-2010FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.Terminated
Jun 13, 2006Z-0745-2007Mispackaging: some kits of peelable sheaths/dilators may contain a 10F sheath instead of a 9F sheath--(Recalling Firm was notified by supplier B. Braun Medical of the error)Terminated
Jun 13, 2006Z-0746-2007Mispackaging: some kits of peelable sheaths/dilators may contain a 10F sheath instead of a 9F sheath--(Recalling Firm was notified by their supplier B. Braun Medical of the error)Terminated
Oct 28, 2005Z-0340-06Device kits contain B. Braun Locking Tearaway Introducer Systems labeled as 7Fr but may contain 8 Fr sheaths.Terminated
Jul 22, 2005Z-1476-05, Z-1481-05The Tearaway Introducer Sheath may have cracked hubs/handles and/or improper peeling during use.Terminated

Other LJT clearances

The newest 510(k)s for the same product code; their median review was 86 days; see every LJT device

Record

Type
Traditional
Received
Sep 5, 1995
Review panel
General Hospital
Regulation
21 CFR 880.5965

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files