Boehringer Mannheim/Hitachi 917 Analyzer

Analyzer, Chemistry (Photometric, Discrete), For Clinical Use. 510(k) K953239, from Boehringer Mannheim Corp. (Indianapolis, IN). Product code JJE, class 1.

As of Nov 18, 2026, K953239 is an FDA 510(k) held by Boehringer Mannheim Corp., substantially equivalent on Sep 25, 1995 after 76 days, with 18 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 25, 1995
  • Review time: 76 days
  • Recalls: 18
  • Cited as a predicate by: 6 devices

Cited as a Predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K953239, newest first; each opens FDA's recall record
Dec 18, 2017Z-1269-2018Complaints have been received concerning questionable results on analyzers due to customers utilizing sample 13 mm tubes without tube adapters.Terminated
Jun 29, 2017Z-0378-2018Roche has confirmed that a possible sample mismatch issue may occur on the MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer due to a software limitation.Terminated
Jul 15, 2016Z-0094-2017Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning…Terminated
May 7, 2015Z-1922-2015Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid…Terminated
Nov 12, 2008Z-1066-2009If a masked module is unmasked during the reagent registration process, the calibration parameters may be assigned to the wrong reagents, resulting in the reporting of incorrect control and sample results.Terminated
Nov 12, 2008Z-1067-2009If a masked module is unmasked during the reagent registration process, the calibration parameters may be assigned to the wrong reagents, resulting in the reporting of incorrect control and sample results.Terminated
Nov 12, 2008Z-1065-2009If a masked module is unmasked during the reagent registration process, the calibration parameters may be assigned to the wrong reagents, resulting in the reporting of incorrect control and sample results.Terminated
Nov 12, 2008Z-1068-2009If a masked module is unmasked during the reagent registration process, the calibration parameters may be assigned to the wrong reagents, resulting in the reporting of incorrect control and sample results.Terminated
Mar 21, 2008Z-0866-2008Possible mismatch between patient and result.Terminated
Mar 21, 2008Z-0870-2008Possible mismatch between patient and result.Terminated
Mar 21, 2008Z-0869-2008Possible mismatch between patient and result.Terminated
Mar 21, 2008Z-0867-2008Possible mismatch between patient and result.Terminated
Jul 21, 2006Z-1447-06Expired product (dated 2006.06) was shipped as replacement for recalled product.Terminated
Jul 14, 2006Z-1352-06A design change in the electrodes will result in incorrect potassium level results.Terminated
Aug 6, 2004Z-1430-04A software defect will allow the reporting out of believable, but clinically significant, erroneous results under certain conditions.Terminated
Aug 6, 2004Z-1429-04A software defect will allow the reporting out of believable, but clinically significant, erroneous results under certain conditions.Terminated
Aug 6, 2004Z-1431-04A software defect will allow the reporting out of believable, but clinically significant, erroneous results under certain conditions.Terminated
May 8, 2003Z-0012-04Incorrect results, without a data flag, reported if one of the other reagent containers becomes depleted before the R1 reagent is depleted.Terminated

Record

Type
Traditional
Received
Jul 11, 1995
Review panel
Clinical Chemistry
Regulation
21 CFR 862.2160

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files