Lipase Flex Reagent Cartridge

Lipase-Esterase, Enzymatic, Photometric, Lipase. 510(k) K952816, from Dupont Medical Products (Wilmington, DE). Product code CHI, class 1.

As of Nov 28, 2026, K952816 is an FDA 510(k) held by Dupont Medical Products, substantially equivalent on Jul 20, 1995 after 30 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 30 days
  • Recalls: 3

Recalls

Recalls FDA ties to K952816, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 7, 2018Z-0465-2019, Z-0466-2019There is lot to lot variability for QC and patient results when using the identified reagent cartridge lots.Terminated
Mar 10, 2010Z-0588-2011Open Well instability: After calibration, QC shifts high (out of range) after the Flex Well is hydrated. Shifts of up to 20% have been observed in both QC and patient samplesas soon as two hours post hydration.Terminated

Record

Type
Traditional
Received
Jun 20, 1995
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1465

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.