Arca
Skip to content
Tools
Tools

Darwin Knee System

Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer. 510(k) K943462, from Johnson & Johnson Professionals, Inc. (Raynham, MA). Product code JWH, class 2.

As of Nov 20, 2026, K943462 is an FDA 510(k) held by Johnson & Johnson Professionals, Inc., substantially equivalent on Dec 21, 1994 after 156 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 156 days
  • Recalls: 8
  • Cited as a predicate by: 12 devices

Cited as a predicate

12 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K943462, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 9, 2013Z-1077-2014Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength…Terminated
May 7, 2009Z-1686-2009There may be a crack on the lateral side of the condyle in the posterior chamfer region. The device is used as a Orthopedic knee implant.Terminated
Sep 27, 2007Z-0405-2008, Z-0404-2008, Z-0406-2008, Z-0402-2008, Z-0407-2008, Z-0403-2008An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.Terminated

Other JWH clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every JWH device

Record

Type
Traditional
Received
Jul 18, 1994
Review panel
Orthopedic
Regulation
21 CFR 888.3560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files