Steri-Vac Gas Sterilizer Model 8XL
Sterilizer, Ethylene-Oxide Gas. 510(k) K941748, from 3M Health Care, Ltd. (St. Paul, MN). Product code FLF, class 2.
As of Nov 22, 2026, K941748 is an FDA 510(k) held by 3M Health Care, Ltd., substantially equivalent on Dec 2, 1994 after 238 days, with 3 recalls tied to it in FDA's recall database.
- Review time: 238 days
- Recalls: 3
- Cited as a predicate by: 1 device
- 3M Steri-Vac Sterilizer/AeratorK142034 · 3M Company · Jan 8, 2015
| Jan 22, 2016 | Z-1563-20163M discovered a valve with incorrectly labeled "input" and "output" orientation. This defect can result in the valve being installed backwards, creating a remote possibility of valve failure. A failure may result in EO… | Terminated |
| Mar 17, 2014 | Z-1318-2014Recently, 3M received a single report of a potential exposure to Ethylene Oxide (EO) at a hospital installation involving 3M" Steri-Vac" Gas Sterilizers. The cause was traced to a rare situation involving a failure of… | Terminated |
| Sep 13, 2012 | Z-0022-20133M determined that the electrical grounding mechanism for the Steri-Vac 5XL and 8XL gas Sterilizer/Aerator did not meet all of the requirements of the standard in that the electrical wiring, which is insulated and… | Terminated |
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 8, 1994
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.6860
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files