Cruciate Retaining Cemented Tibial Components

Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer. 510(k) K932776, from Exactech, Inc. (Gainesville, FL). Product code JWH, class 2.

As of Nov 19, 2026, K932776 is an FDA 510(k) held by Exactech, Inc., substantially equivalent on Oct 20, 1994 after 499 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 499 days
  • Recalls: 2
  • Cited as a predicate by: 3 devices

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K932776, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 19, 2005Z-0927-06An implant within a lot of finned tibial trays manufactured at Exactech was not made to specification. The locking cavity in the tibial tray was not fully machined thus preventing the tibial insert from locking in place.Terminated
Jun 2, 2005Z-1379-05Implant was identified incorrectly at the time of manufacture and then laser etched with the incorrect size information.Terminated

Other JWH clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every JWH device

Record

Type
Traditional
Received
Jun 8, 1993
Review panel
Orthopedic
Regulation
21 CFR 888.3560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files