Surgical Sutures, Pgl, Gut, Deklene, Opthamend Etc
Suture, Absorbable, Natural. 510(k) K930738, from Deknatel, Inc. (San Jose-Heredia, CR). Product code GAL, class 2.
As of Nov 20, 2026, K930738 is an FDA 510(k) held by Deknatel, Inc., substantially equivalent on Jul 26, 1994 after 529 days, with 9 recalls tied to it in FDA's recall database.
- Review time: 529 days
- Recalls: 9
- Cited as a predicate by: 9 devices
- GDC-10 Stretch Resistant Guglielmi Detachable Coil, GDC-10 2-Diameter Stretch Resistant Guglielmi Detachable Coil, GDC-1K993415 · Boston Scientific, Target · Jan 21, 2000
- Cutalon Nylon Polyamide Surgical SutureK981322 · Genzyme Surgical Product (Genzyme Biosurgery) · Jul 7, 1998
- Deklene MAXXK193530 · Teleflex Medical · Mar 19, 2020
- silky II POLYDEK, cottony II, TEVDEK II, NextStitch, DEKLENE II, DEKLENE MAXX Gabbay-Frater, NYLON, SILK, STAINLESS…K153076 · Teleflexmedical, Inc. · Jun 16, 2016
- Digitex Suture Delivery System, Digitexpolypropylene Suture Cartridge Size 0, With Needle and Without Needle, DigitexK120284 · Coloplast A/S · May 23, 2012
- Siesta Medical, Inc., Prelude Tongue Suspension SystemK101060 · Siesta Medical, Inc. · Jul 8, 2010
- Force Fiber Black Co-Braid Polyethylene Non-Absorbable Surgical SutureK070673 · Teleflexmedical, Inc. · Apr 2, 2007
- Lsi Solutions Suture Quick Load ProductsK031443 · Lsi Solutions · Jun 18, 2003
- Cottony Ii, Silky Ii Polydek & Tevdek Ii Polyester SutureK021019 · Genzyme Corp. · Jun 18, 2002
| Mar 11, 2014 | Z-1557-2014, Z-1561-2014, Z-1559-2014, Z-1560-2014, Z-1556-2014, Z-1555-2014, Z-1558-2014Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use. | Terminated |
| Mar 11, 2014 | Z-1497-2014The products are being recalled because they did not meet minimum diameter requirements. | Terminated |
| Jun 6, 2007 | Z-1018-2007Suture Diameter/Strength specification failure: The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage. | Terminated |
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Feb 12, 1993
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4830
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files