DFC-1024 & DFC-512 Digital Imaging System
Camera, Ophthalmic, Ac-Powered. 510(k) K913929, from Ophthalmic Imaging Systems (Sacramento, CA). Product code HKI, class 2.
As of Nov 19, 2026, K913929 is an FDA 510(k) held by Ophthalmic Imaging Systems, substantially equivalent on Nov 27, 1991 after 85 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 85 days
- Recalls: 5
- Cited as a predicate by: 4 devices
- Ophthavision Imaging SystemK980295 · Mrp Group, Inc. · May 19, 1998
- Discam Digital Imaging SystemK974535 · Marcher Enterprises, Ltd. · Aug 14, 1998
- Imagescape Digital Retinal Image SystemK971685 · Tomey Corporation USA · Jun 24, 1997
- Ois Eyescan Portable Modular Imaging System, Model Ois EyescanK092374 · Ophthalmic Imaging Systems · Nov 17, 2009
| Dec 9, 2016 | Z-1017-2017This recall has been initiated due to an issue related to the potential accidental deletion of record(s) by an Eye Station user. | Terminated |
| Dec 9, 2016 | Z-1142-2017Merge received reports of Merge Eye Station being unable to capture images of the eye as expected per the intended use of the product. Merge is recalling product from the field to reduce the risk to patient health. | Terminated |
| Apr 4, 2016 | Z-1828-2017During an antivirus program scan of the Eye Station or Eye Care PACS, the antivirus program may detect a newly captured image to be an unwanted file and delete the image permanently. | Terminated |
| Apr 4, 2016 | Z-1681-2017The system will completely stop allowing the downloading or acquisition of new images due to incompatible software application versions. | Terminated |
| Mar 28, 2016 | Z-1246-2017Eye Station images were not importing properly and were imported under "unknown" due to an issue when validating patients using only an Medical Record Number (MRN) | Terminated |
- TearViewK253176 · Beyond 700 Pty, Ltd. · Jun 22, 2026
- Eyer 2K251353 · Phelcom Technologies · Jan 16, 2026
- Fundus On Phone Non Mydriatic (FOP NM-10)K252120 · Remidio Innovative Solutions Private Limited · Dec 12, 2025
- Optina-4C (MHRC-C1N)K250770 · Optina Diagnostics, Inc. · Sep 4, 2025
- 3nethra neo HD FA; 3nethra neo HDK243600 · Forus Health Pvt.Ltd · Aug 14, 2025
- Resolve Fundus CameraK250683 · Optain Health, Inc. · Apr 30, 2025
- Sentinel CameraK243664 · Ai Optics · Dec 17, 2024
- Verily Numetric Retinal CameraK242508 · Verily Life Sciences, LLC · Dec 9, 2024
- Type
- Traditional
- Received
- Sep 3, 1991
- Review panel
- Ophthalmic
- Regulation
- 21 CFR 886.1120
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files