Arca
Skip to content
Tools
Tools

Varian C-Series Clinacs (Clinac 2300C/D)

Accelerator, Linear, Medical. 510(k) K913119, from Varian Assoc., Inc. (Palo Alto, CA). Product code IYE, class 2.

As of Nov 20, 2026, K913119 is an FDA 510(k) held by Varian Assoc., Inc., substantially equivalent on Nov 7, 1991 after 115 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 115 days
  • Recalls: 6
  • Cited as a predicate by: 2 devices

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K913119, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 23, 2010Z-1249-2011Position errors may result in misdelivery of radiation.Terminated
Jun 10, 2010Z-0886-2011The Auto goto or Auto setup functions in C-Series version 7 software ignore the couch angle and exceed motion zone (unless a tolerance is defined by the user); it has the potential for collision with the patient on the…Terminated
Oct 8, 2009Z-0910-2011Product may deliver radiation treatment to areas larger than intended,to healthy tissue.Terminated
May 2, 2007Z-1074-2007The wedge tray may slip out of the Clinac interface mount during gantry rotation.Terminated
Aug 31, 2006Z-0194-2007The chain holding the device gantry in position may break.Terminated
Mar 30, 2005Z-1530-05Emergency Off Switch located on the device may not function correctly resulting in the device remaining powered on after the switch is activated.Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Jul 15, 1991
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.