Stereotactic Irradiation System

Accelerator, Linear, Medical. 510(k) K904908, from Northwest Medical Physics Center (Lynnwood, WA). Product code IYE, class 2.

As of Nov 25, 2026, K904908 is an FDA 510(k) held by Northwest Medical Physics Center, substantially equivalent on Jun 6, 1991 after 218 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 218 days
  • Recalls: 8

Recalls

Recalls FDA ties to K904908, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 18, 2008Z-2137-2008A localization may be produced which has the incorrect moves, and therefore, could place the patient in the incorrect location.Terminated
Apr 7, 2006Z-1146-2007This correction was for several software bugs which caused image format problems; incorrect ruler display; software crashes; failure to update pixel spacing; display of unattainable treatment table positions; failure to…Terminated
Jan 2, 2006Z-1145-2007Isoloc version 6.5 and 4.5 software issued a false warning message regarding Gantry angles when set to "Anatomic Landmark" setting.Terminated
May 31, 2005Z-1147-2007When the ISOLOC (Version 4.2) profile distance unit is set to "centimeter" and the user clicked the "back" button in the digitizer window, the marker and target coordinates in the localization window erroneously display…Terminated
Apr 28, 2005Z-1152-2007When ISOLOC version 6.0 and 4.2 distance unit is set to centimeter, the localization coordinates were rounded to the nearest millimeter which could affect positioning by 0.5 mm.Terminated
Apr 11, 2005Z-1153-2007When ISOLOC version 6.0 was used in the centimeter mode, the display in the couch moves window had a rounding error relative to the Portal Image Marker window of up to 0.5 mm.Terminated
Dec 30, 2004Z-1154-2007When ISOLOC version 6.0 software was used in the Anatomic Landmark mode, the couch moves display in the Couch Translator window was incorrect.Terminated
Nov 30, 2004Z-1155-2007When ISOLOC Version 4.2 unit is set to centimeter and the user clicks the Reenter button in the digitizer window, the coordinates of the markers and target in the anatomic imaging system erroneously displays values ten…Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Oct 31, 1990
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files