Arca
Skip to content

Leos Laser and Endoscopy System

Laser, Ophthalmic. 510(k) K240615, from Beaver-Visitec International, Inc. (Waltham, MA). Product code HQF, class 2.

As of Nov 17, 2026, K240615 is an FDA 510(k) held by Beaver-Visitec International, Inc., substantially equivalent on Apr 3, 2025 after 394 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Apr 3, 2025
  • Review time: 394 days
  • Recalls: 1

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Leos system is indicated for intraoperative photocoagulation of the ciliary processes in the treatment of glaucoma, proliferative retinopathies, retinal detachment, and for evaluation of the internal ocular structures in patients with dense opacifications of the anterior segment which do not allow a posterior view.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K240615, newest first; each opens FDA's recall record
StartedReasonStatus
Dec 5, 2025Z-1008-2026Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB…Open, Classified

Other HQF Clearances

The newest 510(k)s for the same product code; their median review was 264 days; see every HQF device

Record

Type
Traditional
Received
Mar 5, 2024
Review panel
Ophthalmic
Regulation
21 CFR 886.4390

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.