Arca
Skip to content
Tools
Tools

Hudson RCI Dri-Tech Breathing Circuits and 6 more

Heater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer. 510(k) K234032, from Medline Industries, LP (Northfiled, IL). Product code BZE, class 2.

As of Nov 18, 2026, K234032 is an FDA 510(k) held by Medline Industries, LP, substantially equivalent on Jun 4, 2024 after 167 days, with no recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jun 4, 2024
  • Review time: 167 days
  • Recalls: 0

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Hudson RCI Dri-Tech Breathing Circuits are intended for pediatric and adult patients in professional healthcare environments as a conduit of breathing gas between a patient and a ventilator and to maintain the temperature of humidified inspired gas.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Other BZE Clearances

The newest 510(k)s for the same product code; their median review was 268 days; see every BZE device

Record

Type
Traditional
Received
Dec 20, 2023
Review panel
Anesthesiology
Regulation
21 CFR 868.5270

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.