Arca
Skip to content
Tools
Tools

StoneSmart Connect Console, LithoVue Elite Standard Deflection, LithoVue Elite Reverse Deflection

Ureteroscope And Accessories, Flexible/Rigid. 510(k) K233645, from Boston Scientific Corporation - (Marlborough, MA). Product code FGB, class 2.

As of Nov 21, 2026, K233645 is an FDA 510(k) held by Boston Scientific Corporation -, substantially equivalent on Dec 12, 2023 after 28 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 28 days
  • Recalls: 1
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The LithoVue Elite Digital Flexible Ureteroscope System is intended to by used to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral or percutaneous access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K233645, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Jul 22, 2026Z-3188-2026A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during…Open

Other FGB clearances

The newest 510(k)s for the same product code; their median review was 140 days; see every FGB device

Record

Type
Special
Received
Nov 14, 2023
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1500

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files