Arca
Skip to content
Tools
Tools

FARADRIVE™ Steerable Sheath

Catheter, Steerable. 510(k) K233248, from Boston Scientific Corporation (St Paul, MN). Product code DRA, class 2.

As of Nov 18, 2026, K233248 is an FDA 510(k) held by Boston Scientific Corporation, substantially equivalent on Dec 11, 2023 after 74 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Dec 11, 2023
  • Review time: 74 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The FARADRIVE Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart, including the left side of the heart through the interatrial septum.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K233248, newest first; each opens FDA's recall record
Jun 11, 2026Z-2757-2026Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve…Open, Classified
Jun 11, 2026Z-2756-2026Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve…Open, Classified
Jun 11, 2026Z-2755-2026Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve…Open, Classified

Other DRA Clearances

The newest 510(k)s for the same product code; their median review was 89 days; see every DRA device

Record

Type
Traditional
Received
Sep 28, 2023
Review panel
Cardiovascular
Regulation
21 CFR 870.1280

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.