FARADRIVE™ Steerable Sheath
Catheter, Steerable. 510(k) K233248, from Boston Scientific Corporation (St Paul, MN). Product code DRA, class 2.
As of Nov 18, 2026, K233248 is an FDA 510(k) held by Boston Scientific Corporation, substantially equivalent on Dec 11, 2023 after 74 days, with 3 recalls tied to it in FDA's recall database.
- Substantially equivalent: Dec 11, 2023
- Review time: 74 days
- Recalls: 3
- Cited as a predicate by: 2 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe FARADRIVE Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart, including the left side of the heart through the interatrial septum.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- POLARSHEATH™ Steerable Sheath 12F (UPN:M004CRBS3150), POLARMAP™ Circular Mapping Catheter (UPN:M004CRBS7210), POLARMAP™…K223824 · Boston Scientific Corporation · Mar 22, 2023
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- FARADRIVE™ Steerable Sheath (M004PF21M402)K262165 · Boston Scientific Corporation · Jul 24, 2026
- Globe Introducer (601-01001)K250529 · Kardium, Inc. · Jun 3, 2025
Recalls
Recalls FDA ties to K233248, newest first; each opens FDA's recall record| Jun 11, 2026 | Z-2757-2026Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve… | Open, Classified |
| Jun 11, 2026 | Z-2756-2026Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve… | Open, Classified |
| Jun 11, 2026 | Z-2755-2026Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve… | Open, Classified |
Other DRA Clearances
The newest 510(k)s for the same product code; their median review was 89 days; see every DRA device- FARADRIVE™ Steerable Sheath (M004PF21M402)K262165 · Boston Scientific Corporation · Jul 24, 2026
- Globe Introducer (601-01001)K250529 · Kardium, Inc. · Jun 3, 2025
- FlexCath Contour™ Steerable SheathK232321 · Medtronic, Inc. · Oct 31, 2023
- POLARSHEATH™ Steerable Sheath 12F (UPN:M004CRBS3150), POLARMAP™ Circular Mapping Catheter (UPN:M004CRBS7210), POLARMAP™…K223824 · Boston Scientific Corporation · Mar 22, 2023
- MitraClip G4 Steerable Guide CatheterK221397 · Abbott Medical · Sep 28, 2022
- AcQGuide® VUE Steerable SheathK221044 · Acutus Medical, Inc. · May 5, 2022
- AcQGuide MAX Steerable SheathK211100 · Acutus Medical, Inc. · May 14, 2021
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 28, 2023
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1280
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files