da Vinci Surgical System Model IS5000, da Vinci Insufflator and Tube Set with Smoke Evacuation

System, Surgical, Computer Controlled Instrument. 510(k) K232610, from Intuitive Surgical, Inc. (Sunnyvale, CA). Product code NAY, class 2.

As of Nov 19, 2026, K232610 is an FDA 510(k) held by Intuitive Surgical, Inc., substantially equivalent on Mar 14, 2024 after 199 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 199 days
  • Recalls: 6
  • Cited as a predicate by: 6 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Model IS5000) shall assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K232610, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 24, 2026Z-2310-2026Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the handheld camera to the robotic…Open
Dec 30, 2025Z-1355-2026Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both…Open
Nov 17, 2025Z-1428-2026An error event following the release of a software version for the surgical system that can result in loss of user interface content on an external monitor or tower monitor.Open
Nov 7, 2025Z-1232-2026Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion…Open
Feb 26, 2025Z-1325-2025Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressedOpen
Aug 21, 2024Z-0964-2026Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be…Open

Other NAY clearances

The newest 510(k)s for the same product code; their median review was 89 days; see every NAY device

Record

Type
Traditional
Received
Aug 28, 2023
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1500

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files