Medline ReNewal Reprocessed LigaSure Exact Dissector, Nano-coated (LF2019)
Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed. 510(k) K232511, from Surgical Instrument Service and Savings, Inc. (Redmond, OR). Product code NUJ, class 2.
As of Nov 17, 2026, K232511 is an FDA 510(k) held by Surgical Instrument Service and Savings, Inc., substantially equivalent on Apr 29, 2024 after 255 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Apr 29, 2024
- Review time: 255 days
- Recalls: 1
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Medline ReNewal Reprocessed LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- LigaSure Exact Dissector, Nano-coatedK202917 · Covidien, LLC · Mar 23, 2021
Recalls
Recalls FDA ties to K232511, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Oct 1, 2025 | Z-0467-2026Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged. | Open, Classified |
Other NUJ Clearances
The newest 510(k)s for the same product code; their median review was 88 days; see every NUJ device- Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider (LF1212A)K230251 · Stryker Sustainability Solutions · Mar 28, 2023
- Reprocessed LigaSure Curved, Small Jaw, Open Sealer/Divider (LF1212A)K220411 · Stryker Sustainability Solutions · May 13, 2022
- Green OR Reprocessed Aquamantys Bipolar SealerK211203 · Green Or, LLC · Apr 7, 2022
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 18, 2023
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4400
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files