Arca
Skip to content
Tools
Tools

Ct 5300

System, X-Ray, Tomography, Computed. 510(k) K232491, from Philips Healthcare (Suzhou) Co., Ltd. (Suzhou, CN). Product code JAK, class 2.

As of Nov 18, 2026, K232491 is an FDA 510(k) held by Philips Healthcare (Suzhou) Co., Ltd., substantially equivalent on May 3, 2024 after 260 days, with 5 recalls tied to it in FDA's recall database.

  • Substantially equivalent: May 3, 2024
  • Review time: 260 days
  • Recalls: 5
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The CT 5300 is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • Philips Incisive CTK212441 · Philips Healthcare (Suzhou) Co., Ltd. · Apr 27, 2022
  • Precise ImageK210760 · Philips Medical Systems Nederland B.V. · Jan 14, 2022
  • Precise PositionK203514 · Philips Healthcare (Suzhou) Co., Ltd. · Jun 17, 2021

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K232491, newest first; each opens FDA's recall record
Oct 29, 2025Z-0881-2026If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems…Open, Classified
Oct 29, 2025Z-0882-2026If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems…Open, Classified
Sep 25, 2025Z-0376-2026Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM…Open, Classified
Sep 25, 2025Z-0375-2026Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM…Open, Classified
Sep 25, 2025Z-0377-2026Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM…Open, Classified

Other JAK Clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device

Elsewhere

Record

Type
Traditional
Received
Aug 17, 2023
Review panel
Radiology
Regulation
21 CFR 892.1750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files