Arca
Skip to content
Tools
Tools

SIGNA Victor

System, Nuclear Magnetic Resonance Imaging. 510(k) K223439, from Ge Healthcare (Tianjin) Company Limited (Tianjin, CN). Product code LNH, class 2.

As of Nov 23, 2026, K223439 is an FDA 510(k) held by Ge Healthcare (Tianjin) Company Limited, substantially equivalent on Feb 13, 2023 after 91 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 91 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The SIGNA Victor is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • SIGNA PrimeK211980 · Ge Healthcare (Tianjin) Company Limited · Jan 16, 2022

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K223439, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Jun 14, 2024Z-2480-2024GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE…Open

Other LNH clearances

The newest 510(k)s for the same product code; their median review was 143 days; see every LNH device

Elsewhere

Record

Type
Traditional
Received
Nov 14, 2022
Review panel
Radiology
Regulation
21 CFR 892.1000

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.