Access Testosterone
Radioimmunoassay, Testosterones And Dihydrotestosterone. 510(k) K223405, from Beckman Coulter, Inc. (Chaska, MN). Product code CDZ, class 1.
As of Nov 20, 2026, K223405 is an FDA 510(k) held by Beckman Coulter, Inc., substantially equivalent on Jan 13, 2023 after 65 days, with 1 recall tied to it in FDA's recall database.
- Review time: 65 days
- Recalls: 1
The Access Testosterone assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of total testosterone levels in human serum and plasma using the Access Immunoassay Systems.
- Access Testosterone Reagents for Use On The Access Immunoassay Analyzer, Models 33560, 33565K001935 · Beckman Coulter, Inc. · Aug 22, 2000
| Started | Recall | Status |
|---|---|---|
| May 5, 2025 | Z-1901-2025Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result… | Open |
- IDS-iSYS Total TestosteroneK252728 · Immunodiagnostic Systems Limited · Apr 15, 2026
- IDS-iSYS Free TestosteroneK240865 · Immunodiagnostic Systems Limited · Oct 23, 2024
- Access SHBGK233480 · Beckman Coulter, Inc. · Feb 8, 2024
- Elecsys Testosterone IIK211685 · Roche Diagnostics · May 6, 2022
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 9, 2022
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1680
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files