Optima Coil System
Device, Neurovascular Embolization. 510(k) K223386, from Balt USA, LLC (Irvine, CA). Product code HCG, class 2.
As of Nov 18, 2026, K223386 is an FDA 510(k) held by Balt USA, LLC, substantially equivalent on Dec 6, 2022 after 29 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Dec 6, 2022
- Review time: 29 days
- Recalls: 1
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Optima Coil System is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Optima Coil System is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for…
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Optima Coil SystemK200030 · Balt USA, LLC · Feb 1, 2020
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- Optima Coil System (OptiBlock Line Extension)K242582 · Balt USA, LLC · Sep 27, 2024
Recalls
Recalls FDA ties to K223386, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Jan 3, 2024 | Z-1041-2024Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the… | Open, Classified |
Other HCG Clearances
The newest 510(k)s for the same product code; their median review was 69 days; see every HCG device- Micrusframe (MFR100202); Micrusframe (MFR100253); Micrusframe (MFR100305); Micrusframe (MFR100356); Micrusframe…K262419 · Cerenovus, Inc. · Aug 14, 2026
- Target Xtra Detachable CoilK261772 · Stryker Neurovscular · Jun 24, 2026
- Numen™ Helia Coil Embolization System; NumenFR™ Detachment SystemK260351 · MicroPort NeuroTech (Shanghai) Co., Ltd. · May 20, 2026
- Optima Coil System (OptiOne Coil System)K254221 · Balt USA, LLC · Jan 27, 2026
- Target Detachable CoilK252694 · Stryker Neurovascular · Dec 11, 2025
- InZone IST Detachment System; IZDS Connecting CableK251832 · Stryker Neurovascular · Oct 17, 2025
- Optima Coil System (Optima Packing Coil System)K252700 · Balt USA, LLC · Sep 26, 2025
- Numen™ Coil Embolization System; NumenFR™ Detachment SystemK242154 · MicroPort NeuroTech (Shanghai) Co., Ltd. · Sep 30, 2024
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Nov 7, 2022
- Review panel
- Neurology
- Regulation
- 21 CFR 882.5950
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files