Bronchoscope (Flexible Or Rigid). 510(k) K223144, from Noah Medical Corp. (San Carlos, CA). Product code EOQ, class 2.

As of Nov 17, 2026, K223144 is an FDA 510(k) held by Noah Medical Corp., substantially equivalent on Mar 1, 2023 after 148 days, with 5 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Mar 1, 2023
  • Review time: 148 days
  • Recalls: 5
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Galaxy System and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The Galaxy System and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K223144, newest first; each opens FDA's recall record
Sep 12, 2025Z-0598-2026Due to Users not having access to the Instructions for Use/User Manual due to it being password protected.Open, Classified
Oct 28, 2024Z-0742-2025Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of…Open, Classified
Oct 25, 2024Z-0679-2025Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive…Open, Classified
Oct 25, 2024Z-0693-2025Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is…Open, Classified
Mar 9, 2024Z-1734-2024Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.Open, Classified

Other EOQ Clearances

The newest 510(k)s for the same product code; their median review was 229 days; see every EOQ device

Record

Type
Traditional
Received
Oct 4, 2022
Review panel
Ear, Nose, Throat
Regulation
21 CFR 874.4680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files