Galaxy System
Bronchoscope (Flexible Or Rigid). 510(k) K223144, from Noah Medical Corp. (San Carlos, CA). Product code EOQ, class 2.
As of Nov 17, 2026, K223144 is an FDA 510(k) held by Noah Medical Corp., substantially equivalent on Mar 1, 2023 after 148 days, with 5 recalls tied to it in FDA's recall database.
- Substantially equivalent: Mar 1, 2023
- Review time: 148 days
- Recalls: 5
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Galaxy System and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The Galaxy System and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Monarch Endoscopy Platform (Monarch Platform)K173760 · Auris Surgical Robotics, Inc. · Mar 22, 2018
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- Ion Endoluminal System (IF1000)K252045 · Intuitive Surgical, Inc. · Sep 23, 2025
Recalls
Recalls FDA ties to K223144, newest first; each opens FDA's recall record| Sep 12, 2025 | Z-0598-2026Due to Users not having access to the Instructions for Use/User Manual due to it being password protected. | Open, Classified |
| Oct 28, 2024 | Z-0742-2025Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of… | Open, Classified |
| Oct 25, 2024 | Z-0679-2025Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive… | Open, Classified |
| Oct 25, 2024 | Z-0693-2025Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is… | Open, Classified |
| Mar 9, 2024 | Z-1734-2024Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope. | Open, Classified |
Other EOQ Clearances
The newest 510(k)s for the same product code; their median review was 229 days; see every EOQ device- Single-use Flexible BronchoscopeK261637 · Hunan Vathin Medical Instrument Co., Ltd. · Sep 23, 2026
- Ambu® aScope 5 Broncho 2.7/1.2; Ambu® aScope 5 Broncho 4.2/2.2; Ambu® aScope 5 Broncho…K260309 · Ambu A/S · Sep 16, 2026
- 3.8mm Slim ProVu Video Bronchoscope (090-02-0038U); 5.0mm Regular ProVu Video…K254026 · Flexicare Medical Limited. · Sep 2, 2026
- MONARCH Platform (MON-000008)K260382 · Auris Health, Inc. · Jul 25, 2026
- LungFlow Airway CatheterK254037 · Free Flow Medical, Inc. · Jul 10, 2026
- FUJIFILM Endoscope Model EB-840S; FUJIFILM Endoscope Model EB-840TK253137 · Fujifilm Corporation · Jun 16, 2026
- SmartClip Delivery Catheter (ADR-1715)K260565 · Elucent Medical · May 14, 2026
- Evis Exera Iii Bronchovideoscope (Olympus BF-Q190); Evis Exera Iii Bronchovideoscope…K252341 · Olympus Medical Systems Corp. · Apr 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Oct 4, 2022
- Review panel
- Ear, Nose, Throat
- Regulation
- 21 CFR 874.4680
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files