Arca
Skip to content
Tools
Tools

Solea

Laser, Dental, Soft Tissue. 510(k) K221761, from Convergent Dental, Inc. (Needham, MA). Product code NVK, class 2.

As of Nov 18, 2026, K221761 is an FDA 510(k) held by Convergent Dental, Inc., substantially equivalent on Sep 14, 2022 after 89 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Sep 14, 2022
  • Review time: 89 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Solea system is indicated for the following: - Ablation of hard tissue for caries removal and cavity prevention - Incision, excision, vaporization, coagulation and hemostasis of soft tissue in the oral cavity - Cutting, shaving, contouring and resection of oral osseous tissue (both) - Aiding in the reduction of mineral loss in dental enamel

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • SoleaK151306 · Convergent Dental, Inc. · Sep 23, 2015

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K221761, newest first; each opens FDA's recall record
StartedReasonStatus
Sep 1, 2023Z-0324-2024It has been found that potential unintended laser activation can occur without foot pedal depression.Open, Classified

Other NVK Clearances

The newest 510(k)s for the same product code; their median review was 219 days; see every NVK device

Record

Type
Traditional
Received
Jun 17, 2022
Review panel
General, Plastic Surgery
Regulation
21 CFR 878.4810

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files