Arca
Skip to content

Visera Hysterovideoscope Olympus HYF Type V

Hysteroscope (And Accessories). 510(k) K221557, from Olympus Medical Systems Corporation (Tokyo, JP). Product code HIH, class 2.

As of Nov 17, 2026, K221557 is an FDA 510(k) held by Olympus Medical Systems Corporation, substantially equivalent on Sep 2, 2022 after 94 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 2, 2022
  • Review time: 94 days
  • Recalls: 2

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

VISERA HYSTEROVIDEOSCOPE OLYMPUS HYF TYPE V has been designed to be used with an Olympus video system center, light source, documentation equipment, display monitor, and other ancillary equipment for endoscopic diagnosis within the uterus. Do not use the instrument for any purpose other than its intended use.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K221557, newest first; each opens FDA's recall record
Sep 25, 2023Z-0336-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for…Open, Classified
Jul 20, 2023Z-2480-2023IFU labeling is being updated to revise EtO gas mixture and elimination of reusable cleaning brushes.Open, Classified

Other HIH Clearances

The newest 510(k)s for the same product code; their median review was 168 days; see every HIH device

Record

Type
Traditional
Received
May 31, 2022
Review panel
Obstetrics/Gynecology
Regulation
21 CFR 884.1690

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files