Intervertebral Fusion Device With Integrated Fixation, Lumbar. 510(k) K221545, from Medacta International S.A. (Castel San Pietro (Ch), CH). Product code OVD, class 2.

As of Nov 17, 2026, K221545 is an FDA 510(k) held by Medacta International S.A., substantially equivalent on Oct 24, 2022 after 150 days, with 6 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Oct 24, 2022
  • Review time: 150 days
  • Recalls: 6
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The MectaLIF Anterior is an interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The interior of the spacer component of the MectaLIF Anterior can be packed with autograft or autologous bone graft.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K221545, newest first; each opens FDA's recall record
Mar 8, 2023Z-1403-2023There is a potential for breakage of the small screw used to affix the anti-backout plate.Completed
Mar 8, 2023Z-1401-2023There is a potential for breakage of the small screw used to affix the anti-backout plate.Completed
Mar 8, 2023Z-1404-2023There is a potential for breakage of the small screw used to affix the anti-backout plate.Completed
Mar 8, 2023Z-1400-2023There is a potential for breakage of the small screw used to affix the anti-backout plate.Completed
Mar 8, 2023Z-1402-2023There is a potential for breakage of the small screw used to affix the anti-backout plate.Completed
Mar 8, 2023Z-1405-2023There is a potential for breakage of the small screw used to affix the anti-backout plate.Completed

Other OVD Clearances

The newest 510(k)s for the same product code; their median review was 95 days; see every OVD device

Record

Type
Traditional
Received
May 27, 2022
Review panel
Orthopedic
Regulation
21 CFR 888.3080

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files