Arca
Skip to content
Tools
Tools

Synergy Cranial v2.2.9, StealthStation Cranial v3.1.4

Neurological Stereotaxic Instrument. 510(k) K221087, from Medtronic Navigation, Inc. (Louisville, CO). Product code HAW, class 2.

As of Nov 20, 2026, K221087 is an FDA 510(k) held by Medtronic Navigation, Inc., substantially equivalent on Jun 10, 2022 after 58 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 58 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Synergy Cranial v2.2.9: The StealthStation System, with Synergy Cranial software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical procedures. The system is indicated for any medical condition in which reference to a rigid anatomical structure can be identified relative to images of the anatomy.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K221087, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 11, 2023Z-1666-2023In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain…Open
Apr 11, 2023Z-1630-2023During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information…Open

Other HAW clearances

The newest 510(k)s for the same product code; their median review was 162 days; see every HAW device

Record

Type
Traditional
Received
Apr 13, 2022
Review panel
Neurology
Regulation
21 CFR 882.4560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.