Arca
Skip to content
Tools
Tools

Evidence MultiSTAT DOA Urine MultiPlex, Evidence MultiSTAT

Enzyme Immunoassay, Opiates. 510(k) K220451, from Randox Laboratories Limited (Crumlin, GB). Product code DJG, class 2.

As of Nov 20, 2026, K220451 is an FDA 510(k) held by Randox Laboratories Limited, substantially equivalent on Oct 5, 2023 after 595 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 595 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Evidence MultiSTAT DOA Urine MultiPlex is intended for use with the Evidence MultiSTAT. The Evidence MultiSTAT is an analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine at the associated cutoffs.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K220451, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Apr 5, 2024Z-1786-2024Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This…Open

Other DJG clearances

The newest 510(k)s for the same product code; their median review was 115 days; see every DJG device

Record

Type
Traditional
Received
Feb 17, 2022
Review panel
Clinical Toxicology
Regulation
21 CFR 862.3650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files