Monitor B105M, Monitor B125M, Monitor B155M, Monitor B105P, Monitor B125P
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K213490, from Ge Medical Systems Information Technologies, Inc. (Wauwatosa, WI). Product code MHX, class 2.
As of Nov 17, 2026, K213490 is an FDA 510(k) held by Ge Medical Systems Information Technologies, Inc., substantially equivalent on Apr 1, 2022 after 151 days, with 6 recalls tied to it in FDA's recall database.
- Substantially equivalent: Apr 1, 2022
- Review time: 151 days
- Recalls: 6
- Cited as a predicate by: 2 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe monitor B105M, B125M,B155M, B105P and B125P are portable multi-parameter patient monitors intended to be used for monitoring, recording, and to generate alarms for multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Monitor B125, Monitor B105K201941 · Ge Medical Systems Information Technologies, Inc. · Nov 4, 2020
- Carescape B450K191249 · GE Healthcare Finland Oy · Jan 24, 2020
- VC150 Vital Signs MonitorK133810 · Innokas Yhtyma · Oct 30, 2014
- CARESCAPE ONE and CARESCAPE DOCK FO, CARESCAPE ECG, CARESCAPE TEMP, CARESCAPE PRESK200494 · GE Healthcare · Mar 29, 2020
- Datex-Ohmeda S/5 Neuromuscular Transmission Module, E-Nmt And AccessoriesK051635 · GE Healthcare · Jul 15, 2005
- Datex-Ohmeda S/5 Single-Width Airway Module, E-Minic And AccessoriesK052582 · GE Healthcare · Oct 14, 2005
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- Monitor B105M; Monitor B125M; Monitor B155M; Monitor B105P; Monitor B125PK242562 · Ge Medical Systems Information Technologies, Inc. · Jul 9, 2025
- Portrait VSMK233810 · Ge Medical Systems Information Technologies, Inc. · Apr 25, 2024
Recalls
Recalls FDA ties to K213490, newest first; each opens FDA's recall record| Sep 26, 2025 | Z-0455-2026Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of… | Open, Classified |
| Nov 16, 2022 | Z-0733-2023GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that… | Open, Classified |
| Nov 16, 2022 | Z-0731-2023GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that… | Open, Classified |
| Nov 16, 2022 | Z-0732-2023GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that… | Open, Classified |
| Nov 16, 2022 | Z-0729-2023GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that… | Open, Classified |
| Nov 16, 2022 | Z-0730-2023GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that… | Open, Classified |
Other MHX Clearances
The newest 510(k)s for the same product code; their median review was 226 days; see every MHX device- Radius VSM and AccessoriesK260676 · Masimo Corporation · Oct 14, 2026
- Carescape ONE; Carescape Dock F0; TruSignal ECG MM; TruSignal PRES MM; TruSignal TEMP MM;…K260656 · GE Healthcare Finland Oy · Sep 1, 2026
- Carescape Canvas 1000; Carescape Canvas Smart DisplayK260343 · GE Healthcare Finland Oy · Aug 28, 2026
- Infinity® Acute Care System (IACS) VG8.1K260579 · Draeger Medical Systems, Inc. · Aug 20, 2026
- Telemetry Monitor (iT30, iT50)K251508 · Edan Instruments, Inc. · Aug 6, 2026
- Monitor B105M; Monitor B125M; Monitor B155M; Monitor B105P; Monitor B125PK253277 · Ge Medical Systems Information Technologies, Inc. · Jul 24, 2026
- Radius VSM and AccessoriesK254209 · Masimo Corporation · Jun 4, 2026
- M12 Telemetry SystemK252840 · Global Instrumentation, LLC · May 29, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 1, 2021
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1025
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files