Arca
Skip to content
Tools
Tools

Veriqa

Accelerator, Linear, Medical. 510(k) K213370, from Ptw-Freiburg Physikalisch-Technische-Werkstaetten (Freiburg, DE). Product code IYE, class 2.

As of Nov 19, 2026, K213370 is an FDA 510(k) held by Ptw-Freiburg Physikalisch-Technische-Werkstaetten, substantially equivalent on Mar 30, 2022 after 169 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 169 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

VERIQA is a software package for display, evaluation and digital processing of medical image data sets and treatment plans in radiation oncology. VERIQA software is a tool for evaluation and data management of digital medical images and treatment plan information. It supports the medical imaging modalities CT, MR, PET according to the ACR/NEMA DICOM 3.0 standard and other modalities.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • SunCHECKK170307 · Sun Nuclear Corporation · Oct 25, 2017
  • Mirada RtxK130393 · Mirada Medical, Ltd. · Mar 20, 2013

Recalls

Recalls FDA ties to K213370, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 4, 2026Z-3174-2026When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment…Open
Feb 8, 2024Z-1181-2024Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR…Open

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Oct 12, 2021
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files