Arca
Skip to content
Tools
Tools

syngo.via RT Image Suite

System, Planning, Radiation Therapy Treatment. 510(k) K211379, from Siemens Medical Solutions USA, Inc. (Knoxville, TN). Product code MUJ, class 2.

As of Nov 18, 2026, K211379 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Jul 30, 2021 after 87 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 87 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

syngo.via RT Image Suite is a 3D and 4D image visualization, multimodality manipulation and contouring tool that helps the preparation of treatments such as, but not limited to those performed with radiation (for example, Brachytherapy, Particle Therapy, External Beam Radiation Therapy).

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K211379, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 7, 2022Z-0994-2022Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.Terminated
Aug 16, 2021Z-2445-2021Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning…Open

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Elsewhere

Record

Type
Traditional
Received
May 4, 2021
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.