Smith & Nephew Intramedullary Nail Systems
Rod, Fixation, Intramedullary And Accessories. 510(k) K210980, from Smith & Nephew, Inc. (Cordova, TN). Product code HSB, class 2.
As of Nov 17, 2026, K210980 is an FDA 510(k) held by Smith & Nephew, Inc., substantially equivalent on Nov 14, 2022 after 592 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Nov 14, 2022
- Review time: 592 days
- Recalls: 4
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFIndications for interlocking intramedullary nails include simple long bone f r actures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fr actures; prophylactic nailing of impending pathologic f r actures; reconstruction following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Titanium Intramedullary Nail, Titanium Locking ScrewK981529 · Smith & Nephew, Inc. · Jul 9, 1998
- Trigen Retrograde Femoral, Supracondylar And Tibial NailsK051557 · Smith & Nephew, Inc. · Jun 30, 2005
- Trigen Meta-Nail Retrograde Femoral And Tibial NailsK061019 · Smith & Nephew, Inc. · Jun 6, 2006
- Trigen Adolescent TanK040929 · Smith & Nephew, Inc. · May 25, 2004
- Trigen Low Profile Bone ScrewK111025 · Smith & Nephew, Inc. · Jul 1, 2011
- Trigen Hindfoot Fusion NailK043052 · Smith & Nephew, Inc. · Nov 24, 2004
Recalls
Recalls FDA ties to K210980, newest first; each opens FDA's recall record| Mar 16, 2026 | Z-2456-2026Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent… | Open, Classified |
| Mar 18, 2025 | Z-1806-2025Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and… | Open, Classified |
| Jul 24, 2023 | Z-2444-2023A complaint was received indicating that a package contained a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT implant instead of a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE… | Open, Classified |
| Jul 24, 2023 | Z-2443-2023A complaint was received indicating that a package contained a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT implant instead of a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE… | Open, Classified |
Other HSB Clearances
The newest 510(k)s for the same product code; their median review was 192 days; see every HSB device- Freedom Nail Ankle Fusion SystemK261189 · McGinley Orthopaedic Innovations · Oct 19, 2026
- Pediatric Nailing Platform FemurK261345 · OrthoPediatrics Corp. · Sep 24, 2026
- Pangea Femur Reconstruction SystemK261481 · Stryker GmbH · Jun 29, 2026
- AND Medical Femoral Nail System (PFNA01); AND Medical Femoral Nail System (PFNA02)K253080 · Suzhou and Science & Technology Development Corp. · Jun 18, 2026
- Precice Max System and Precice Ankle Salvage SystemK261651 · NuVasive Specialized Orthopedics · Jun 18, 2026
- Fule Metallic Lockable Intramedullary Nail SystemsK253155 · Beijing Fule Science & Technology Development Co., Ltd. · Jun 17, 2026
- HERA Interlocking Intramedullary Nailing SystemK253046 · ZheJiang Decans Medical Devices Co., Ltd. · Jun 16, 2026
- Fitbone® Transport And Lengthening SystemK260146 · Orthofix Srl · May 19, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 1, 2021
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3020
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files