Arca
Skip to content
Tools
Tools

Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HD

Accelerator, Linear, Medical. 510(k) K210500, from Elekta Limited (Crawley, GB). Product code IYE, class 2.

As of Nov 20, 2026, K210500 is an FDA 510(k) held by Elekta Limited, substantially equivalent on Jun 11, 2021 after 109 days, with 10 recalls tied to it in FDA's recall database.

  • Review time: 109 days
  • Recalls: 10
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Elekta Medical Linear Accelerator (EMLA) is intended to be used for external beam radiation therapy (EBRT) treatments as determined by a licensed medical practitioner. It is intended to assist a licensed medical practitioner in the delivery of EBRT to defined target volumes, while sparing surrounding normal tissue and critical organs from excess radiation.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K210500, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 14, 2025Z-1164-2026Due to a manufacturing issue and electrical grounding of systems.Open
Oct 25, 2023Z-0446-2024, Z-0444-2024, Z-0443-2024, Z-0447-2024, Z-0445-2024, Z-0442-2024Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there…Open
May 5, 2023Z-1665-2023A component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601Open
Sep 28, 2021Z-0115-2022There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.Terminated
Sep 17, 2021Z-0116-2022There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Feb 22, 2021
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.