DermaV Laser System
Powered Laser Surgical Instrument. 510(k) K203788, from Lutronic Corporation (Deogyang-Gu, Goyang-Si, KR). Product code GEX, class 2.
As of Nov 17, 2026, K203788 is an FDA 510(k) held by Lutronic Corporation, substantially equivalent on Mar 26, 2021 after 88 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Mar 26, 2021
- Review time: 88 days
- Recalls: 1
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDF1064 nm The DermaV Laser System is indicated for stable long term or permanent hair reduction and for treatment of PFB (Pseudofolliculitis Barbae) on all skin types, Fitzpatrick I – VI, including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- CLARITY II Laser SystemK183566 · Lutronic Corporation · Jun 14, 2019
- Family of CoolGlide Aesthetic LasersK153671 · Cutera, Inc. · Mar 9, 2016
Recalls
Recalls FDA ties to K203788, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Aug 6, 2026 | Z-3110-2026Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn. | Open, Classified |
Other GEX Clearances
The newest 510(k)s for the same product code; their median review was 90 days; see every GEX device- B`orze Inc Laser Family (Sorren/Adonis)K263063 · Borze, Inc. · Oct 30, 2026
- Diode laser device (DT25-BI, DT212-BI, DT213-BI, DT214-BI, DT215-BI)K262349 · Risu Medical Technology Co., Ltd. · Oct 27, 2026
- Fractional CO2 Laser Therapy System (Z121)K262627 · Sanhe PERFECTLASER Technology Co., Ltd. · Oct 26, 2026
- PicoPeak Picosecond Nd:Yag Laser System (Bella-9)K262701 · Beijing Sea Heart International Science and Technologyco.Ltd · Oct 22, 2026
- Picosecond Laser treatment system (Z180)K262628 · Sanhe PERFECTLASER Technology Co., Ltd. · Oct 20, 2026
- Nd:YAG & Alexandrite Long Pulse Laser Therapy Instrument (LPDuo)K261572 · Keyue (Hangzhou) Medical Information Technology Co., Ltd. · Oct 13, 2026
- CureLight Medical Diode Laser Systems (S3-BF45);CureLight Medical Diode Laser Systems (S2-B30);CureLight Medical Diode…K262346 · Wuhan Phomed Technology Co., Ltd. · Oct 8, 2026
- Diode Laser Hair Removal Machine (QDTM-03)K261692 · Beijing Nubway S&T Co., Ltd. · Oct 7, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 28, 2020
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4810
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files