Trinias
Interventional Fluoroscopic X-Ray System. 510(k) K203535, from Shimadzu Corporation Medical Systems (Kyoto, JP). Product code OWB, class 2.
As of Nov 23, 2026, K203535 is an FDA 510(k) held by Shimadzu Corporation Medical Systems, substantially equivalent on Apr 28, 2021 after 146 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 146 days
- Recalls: 2
- Cited as a predicate by: 1 device
The Trinias is an angiographic X-ray system, which is used for diagnostic imaging and interventional procedures. The Trinias is intended to be used for cardiac angiography, neurovascular angiography, abdominal angiography, peripheral angiography, rotational angiography, multi-purpose angiography and whole body radiographic/fluoroscopic procedures
- TriniasK123508 · Shimadzu Corporation · Feb 26, 2014
- TriniasK221922 · Shimadzu Corporation Medical Systems Division · Jul 28, 2022
| Apr 20, 2026 | Z-2294-2026Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off. | Open |
| Sep 30, 2025 | Z-0347-2027The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation. | Open |
- uAngio AVIVA CXK253726 · Shanghai United Imaging Healthcare Co., Ltd. · Aug 21, 2026
- Extron 3;Extron 6K261413 · DRTECH Corporation · Aug 19, 2026
- Zenition 30K254205 · Philips Medical Systems Nederland B.V. · Aug 7, 2026
- Shield System (Siemens 30x40)K260607 · Radiaction , Ltd. · Jul 31, 2026
- SmartCT (R3.0 (L10))K253670 · Philips Medical Systems B.V. · Jul 10, 2026
- C-beamK252735 · Pulmera, Inc. · May 22, 2026
- Intelligent NRK252503 · Canon, Inc. · Apr 30, 2026
- ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS phenoK254173 · Siemens Medical Solutions USA, Inc. · Apr 24, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 3, 2020
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1650
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files