EsophaCap Swallowable Cellular Retrieval Device (changed from Cell-Mata)
Esophagoscope (Flexible Or Rigid). 510(k) K203450, from Capnostics, LLC (Concord, NC). Product code EOX, class 2.
As of Nov 20, 2026, K203450 is an FDA 510(k) held by Capnostics, LLC, substantially equivalent on May 4, 2021 after 162 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 162 days
- Recalls: 2
The gathering and recovery of cells and cellular material from the mucosa in the esophagus for cytological and histological analyses.
- Cell-Mate(Tm) Mass Cytology-Cellular Retrieval SystK934193 · Biosearch Medical Products, Inc. · Nov 19, 1993
- Cytosponge Cell Collection DeviceK142695 · Covidien, LLC · Nov 26, 2014
- Cytosponge Cell Collection DeviceK152794 · Covidien, LLC · Apr 12, 2016
- Cytosponge Cell Collection DeviceK181020 · Covidien, LLC · Aug 16, 2018
| Started | Recall | Status |
|---|---|---|
| Apr 24, 2024 | Z-1900-2024, Z-1899-2024The EsophaCap sponge may be at increased risk of detaching from the string during removal of the device from the patient, particularly at the lower esophageal sphincter at the gastroesophageal junction or the upper… | Open |
- EndoSign® Cell collection device (ES-CYT-102)K233142 · Cyted Limited · Jan 19, 2024
- PENTAX Medical Video Esophagoscope EE17-J10K223072 · Pentax of America, Inc. · Dec 2, 2022
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 23, 2020
- Review panel
- Ear, Nose, Throat
- Regulation
- 21 CFR 874.4710
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files