Electrosurgical Generator ESG-410, Foot Switches, PK Cutting Forceps
Electrosurgical, Cutting & Coagulation & Accessories. 510(k) K203277, from Olympus Winter & Ibe GmbH (Hamburg, DE). Product code GEI, class 2.
As of Nov 20, 2026, K203277 is an FDA 510(k) held by Olympus Winter & Ibe GmbH, substantially equivalent on Apr 28, 2021 after 173 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 173 days
- Recalls: 4
- Cited as a predicate by: 3 devices
ESG-410: The electrosurgical generator, in conjunction with electrosurgical accessories and ancillary equipment, is intended for cutting and coagulation of tissue in the following medical fields: - Open surgery - Laparoscopic surgery - Endoscopic surgery Only for use by a qualified physician in an adequate medical environment.
- ESG-400, Footswitch Double Pedal, Footswitch Single Pedal, P-CordK141225 · Olympus Winter & Ibe GmbH · Aug 12, 2014
- PK Cutting Forceps for ESG-400 WorkstationK142759 · Gyrus Acmi, Inc. · Jan 13, 2015
- Electrosurgical Generator ESG-410 and Accessories (WA91327U, WA91327W), Accessories: Foot switches (WA91311W, WA91321W)K231777 · Olympus Winter & Ibe GmbH · Aug 18, 2023
- Olympus PK Electrosurgical Instruments, PK J-hook, PK Spatula, PK NeedleK213831 · Gyrus Acmi, Inc. · Jan 4, 2022
- POWERSEAL Curved Jaw Sealer and Divider, Double ActionK212643 · Olympus Surgical Technologies America · Sep 27, 2021
| Feb 25, 2026 | Z-1640-2026Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use. | Open |
| Jul 25, 2025 | Z-2339-2025Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'. | Open |
| Jun 21, 2024 | Z-2415-2024Electrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery | Open |
| Dec 15, 2023 | Z-1167-2024The jaw may fracture prior to the procedure during the inspection instructed per the IFU, or during the procedure. | Open |
- SP Electrocautery Device (SP18 Max);SP Electrocautery Device (SP20 Max)K262720 · Single Pass, Inc. · Oct 26, 2026
- SafeAir Electrosurgical Generator with Smoke Evacuation (0795-001-000);SafeAir Generator 3-Pedal Footswitch…K262386 · Stryker Instruments · Oct 20, 2026
- HARMONIC 1100 Shears (HAR1120); HARMONIC 1100 Shears (HAR1136)K254196 · Ethicon Endo-Surgery, LLC · Sep 17, 2026
- Neptune® SafeAir® 10ft Smoke Evacuation Pencil, Square Coated, 70mm Blade, Push Button Switch (0703-046-100);Neptune®…K261959 · Stryker Instruments · Sep 17, 2026
- LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider, Nano-coatedK261070 · Covidien, LLC · Sep 15, 2026
- Potenza PrimeK254185 · Jeisys Medical Incorporated · Sep 11, 2026
- Ellian Enigma RF DeviceK262364 · Ellian Corp · Sep 2, 2026
- Scarlet Pro (Scarlet Pro)K253859 · VIOL Co., Ltd. · Aug 27, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 6, 2020
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4400
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files