VERITAS Phacoemulsification Console, VERITAS Advanced Fluidics Pack and Advanced Infusion Pack, VERITAS Swivel Handpiece, VERITAS Advanced Foot Pedal, VERITAS Remote Control

Unit, Phacofragmentation. 510(k) K203060, from Johnson & Johnson Surgical Vision, Inc. (Santa Ana, CA). Product code HQC, class 2.

As of Nov 17, 2026, K203060 is an FDA 510(k) held by Johnson & Johnson Surgical Vision, Inc., substantially equivalent on Mar 12, 2021 after 155 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Mar 12, 2021
  • Review time: 155 days
  • Recalls: 3

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The VERITAS(TM) Advanced Infusion Pack or VERITAS(TM) Advanced Fluidics Pack contains the manifold (cassette) and Tubing Assembly and is intended to perform irrigation and aspiration during anterior segment cataract surgery. It is used with the VERITAS(TM) Vision System. The VERITAS(TM) Pack is sterilized using ethylene oxide and is designed for single use only.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K203060, newest first; each opens FDA's recall record
May 30, 2023Z-2247-2023Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the…Open, Classified
Dec 7, 2022Z-1005-2023Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and…Open, Classified
Sep 1, 2022Z-0086-2023As a result of returned product complaints for Veritas Phaco packs related to broken or cracked irrigation luer due to a manufacturing issue.Open, Classified

Other HQC Clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every HQC device

Record

Type
Traditional
Received
Oct 8, 2020
Review panel
Ophthalmic
Regulation
21 CFR 886.4670

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.