Flexible Video Cysto-Urethroscope (C-view)
Cystoscope And Accessories, Flexible/Rigid. 510(k) K202957, from KARL STORZ Endoscopy-America, Inc. (El Segundo, CA). Product code FAJ, class 2.
As of Nov 17, 2026, K202957 is an FDA 510(k) held by KARL STORZ Endoscopy-America, Inc., substantially equivalent on Oct 29, 2020 after 29 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: Oct 29, 2020
- Review time: 29 days
- Recalls: 2
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Flexible Video Cysto-Urethroscope C-VIEW is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of urinary tract including the urethra, bladder, ureters, and kidneys.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Flexible HD Cysto-Urethroscope SystemK191357 · KARL STORZ Endoscopy-America, Inc. · Sep 18, 2019
- Flexible Cmos-Video-Cysto-UrethroscopeK131364 · KARL STORZ Endoscopy-America, Inc. · Jan 2, 2014
- Flexible HD Cysto-Urethroscope SystemK182723 · KARL STORZ Endoscopy-America, Inc. · Apr 23, 2019
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- KARL STORZ Flexible HD Cysto-Urethroscope SystemK212476 · Karl Storz Endoscopy America, Inc. · Sep 3, 2021
Recalls
Recalls FDA ties to K202957, newest first; each opens FDA's recall record| May 3, 2022 | Z-1234-2022The endoscope may experience a loss of image when activating monopolar electrocautery, due a manufacturing assembly process issue. | Open, Classified |
| Apr 1, 2022 | Z-1065-2022Failure to achieve the expected six-log reduction in microorganisms following the disinfection process. | Open, Classified |
Other FAJ Clearances
The newest 510(k)s for the same product code; their median review was 123 days; see every FAJ device- Strauss Surgical Cystoscope/Hysteroscope and Accessories InstrumentsK263075 · Strauss Surgical USA · Oct 30, 2026
- OES ELITE Cystoscope and Accessories; Obturator, 17 Fr. (WA2C17BO); Obturator, 19.8 Fr.…K260649 · Olympus Winter & Ibe GmbH · May 21, 2026
- CFlex Cystoscope - Standard Deflection (0570-0455); CFlex Cystoscope - Reverse Deflection…K253696 · Verathon Medical (Canada) Ulc · May 11, 2026
- Strauss Surgical Cystoscopes & Hysteroscopes (STS-2040-302ASA HD-endoscope, 4 x 302 mm,…K253892 · Strauss Surgical USA · Apr 16, 2026
- PrimeSight UltraView SystemK253905 · Cogentix Medical, Inc. · Apr 7, 2026
- Strauss Surgical cystoscope, hysteroscope and accessories InstrumentsK260271 · American Medical Endoscopy, Inc. · Mar 27, 2026
- UV5000W HandleK253226 · Uroviu Corporation · Oct 23, 2025
- Zenflow Spring ScopeK251140 · Zenflow, Inc. · Sep 23, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Sep 30, 2020
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.1500
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files