Arca
Skip to content

LinkSymphoKnee System

Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer. 510(k) K202924, from Waldemar Link GmbH & Co. KG (Norderstedt, DE). Product code JWH, class 2.

As of Nov 17, 2026, K202924 is an FDA 510(k) held by Waldemar Link GmbH & Co. KG, substantially equivalent on May 13, 2021 after 226 days, with 11 recalls tied to it in FDA's recall database.

  • Substantially equivalent: May 13, 2021
  • Review time: 226 days
  • Recalls: 11
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

G eneral : The L in kS ympho K nee S ystem is intended for primary and revision total k nee replacement in s k eletally mature patients with the following conditions. This device is intended for cemented use only unless a cementless modular stem is indicated for use.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K202924, newest first; each opens FDA's recall record
Jul 31, 2025Z-2541-2025Some modular LinkSymphoKnee TiNbN Tibial Components were manufactured with two uncoated security screws, which is incorrect.Open, Classified
Apr 3, 2025Z-1699-2025The device was delivered with a preassembled Femoral Augment screw that was missing its thread.Open, Classified
Apr 3, 2025Z-1700-2025The device was delivered with a preassembled Femoral Augment screw that was missing its thread.Open, Classified
Apr 3, 2025Z-1701-2025The device was delivered with a preassembled Femoral Augment screw that was missing its thread.Open, Classified
Apr 3, 2025Z-1702-2025The device was delivered with a preassembled Femoral Augment screw that was missing its thread.Open, Classified
Feb 14, 2025Z-1372-2025A problem was discovered during the scanning process of the outer carton label because of an incorrect GTIN on the label, which is a GTIN for a different product.Open, Classified
Feb 14, 2025Z-1374-2025A problem was discovered during the scanning process of the outer carton label because of an incorrect GTIN on the label, which is a GTIN for a different product.Open, Classified
Feb 14, 2025Z-1373-2025A problem was discovered during the scanning process of the outer carton label because of an incorrect GTIN on the label, which is a GTIN for a different product.Open, Classified
Apr 25, 2023Z-1674-2023Contains the incorrect implant in the packaging. The packaging incorrectly contains the opposite side of the implant (Lateral-Right/Medial-Left instead of Medial-Right/Lateral-Left).Open, Classified
Dec 23, 2022Z-2066-2023Fixation screw was inserted through the wrong side of the augment (distal to proximal, as opposed to proximal to distal, which is the correct orientation).Open, Classified
Dec 23, 2022Z-2067-2023Fixation screw was inserted through the wrong side of the augment (distal to proximal, as opposed to proximal to distal, which is the correct orientation).Open, Classified

Other JWH Clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every JWH device

Record

Type
Traditional
Received
Sep 29, 2020
Review panel
Orthopedic
Regulation
21 CFR 888.3560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files