Cytrellis Dermal Micro-Coring System
Powered Microneedle Device. 510(k) K202517, from Cytrellis Biosystems, Inc. (Woburn, MA). Product code QAI, class 2.
As of Nov 21, 2026, K202517 is an FDA 510(k) held by Cytrellis Biosystems, Inc., substantially equivalent on Jul 9, 2021 after 311 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 311 days
- Recalls: 2
- Cited as a predicate by: 2 devices
The Cytrellis Dermal Micro-Coring System is indicated for use by medical professionals for the treatment of moderate and severe wrinkles in the mid and lower face in adults aged 22 years or older with Fitzpatrick skin types I-IV.
- Exceed Microneedling DeviceK180778 · Mt. Derm GmbH · Sep 7, 2018
- ellacor System with Micro-Coring TechnologyK252752 · Cytrellis Biosystems, Inc. · Oct 30, 2025
- AI.ME SystemK221011 · Venus Concept USA, Inc. · Dec 20, 2022
| Jun 23, 2023 | Z-2393-2023Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to… | Open |
| Apr 7, 2023 | Z-1648-2023Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring. | Completed |
- Automatic Micro Needle System (CODE-X)K252591 · Woorhi Mechatronics Co., Ltd. · May 13, 2026
- SkinStylus SteriLock MicroSystem (MP1209SL)K253002 · Esthetic Medical, Inc. · Feb 19, 2026
- ellacor System with Micro-Coring TechnologyK252752 · Cytrellis Biosystems, Inc. · Oct 30, 2025
- SER Pen Carain MicroSystem (MP1209SP)K253153 · Su-Ko Technologies, LLC · Oct 29, 2025
- PRO Pen Microneedling System (6883)K243800 · Dermalogica, LLC · Aug 25, 2025
- E-Pen (E-Pen)K243143 · Bomtech Electronics Co., Ltd. · Apr 7, 2025
- 1NEED ProK243472 · Campomats S.R.L. · Mar 19, 2025
- XCELLARISPRO TWIST microneedling deviceK241790 · Dermaroller GmbH · Mar 6, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 1, 2020
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4430
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files