SpaceOAR System

Hydrogel Spacer. 510(k) K202224, from Boston Scientific Corporation (Marlboro, MA). Product code OVB, class 2.

As of Nov 29, 2026, K202224 is an FDA 510(k) held by Boston Scientific Corporation, substantially equivalent on Aug 28, 2020 after 21 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 21 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

SpaceOAR System is intended to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer and in creating this space it is the intent of SpaceOAR System to reduce the radiation dose delivered to the anterior rectum.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K202224, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Feb 24, 2022Z-1002-2022Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct…Open

Other OVB clearances

The newest 510(k)s for the same product code; their median review was 83 days; see every OVB device

Record

Type
Special
Received
Aug 7, 2020
Review panel
Radiology
Regulation
21 CFR 892.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc